Flammability UL 94 V

This method is used to determine the UL 94 V-0, V-1 and V-2 flammability ratings.

It evaluates both the burning and afterglow times after repeated flame application and dripping of the burning test specimen.

Pre-treatment:
2 days / 23 °C / 50% relative humidity
7 days / 70 °C / hot air oven

Flame application 20 mm high Tirill burner flame
Flame application time 2 x 10 s
The second flame application time begins as soon as the ignited specimen is extinguished or immediately if the specimen does not ignite. 

 

Flammability rating UL 94   
  V-0 V-1 V-2
Burning time after flame application (s) ≤10 ≤30 ≤30
Total burning time (s) (10 flame applications) ≤50 ≤250 ≤250
Burning and afterglow times of specimens after second flame application (s) ≤30 ≤60 ≤60
Dripping of burning specimens
(ignition of cotton batting)
no no yes
Specimens completely burned no no no